one. Huge and tiny warehouses can not replenish fuel

Factors: ① The air hole in the glove box foremost to the massive and tiny warehouse may be blocked by plastic film, etc . ② No matter whether the solenoid valve is not open up.

acrylic glovebox : ① Determine the situation of the air gap and clear the pipeline in accordance to the pipeline direction ② Check out the circuit and solenoid valve.

2. The big and little warehouses cannot be pumped

Causes: ① The suction holes in the huge and modest bins are blocked by plastic movie, and so forth. or the door is leaking ② The bin doorways are not closed tightly, or the seal ring of the big and modest bins is free.

Technique: ① Determine the position of the air gap and distinct the pipeline in accordance to the path of the pipeline ② Close the warehouse door and verify the size of the sealing ring ③ Examine no matter whether the functioning air strain is also low (because the big warehouse exhaust valve is a pneumatic valve).

three. The box force carries on to boost for the duration of cleansing

Cause: The cleaning valve is not opened, the intake pressure is way too substantial, and the pipeline is not sleek.

Methods: ①Ensure that the pipeline is easy, and it is very best not to bend it ②When cleaning, the relative stress in the tank is repeatedly under .five mbar.

4. glovebox chemistry of the purification column is faulty



Causes: ① The operating fuel pressure is less than .5Mpa ② The principal valve sign comments failure.

Techniques: ①Replace the functioning gasoline cylinder, or verify the force of the auxiliary gauge of the strain reducing valve at .5Mpa

five. The cycle can not be opened

Reason: Minimal-voltage or higher-voltage safety of admirer inverter.

laboratory glovebox : Following the voltage is steady, shut down and restart following 1 minute.