Glove box h2o probe cleaning (only for washable probes)

one) Cause: If the water probe is not cleaned for a prolonged time, some dirt will adhere to the surface area, impacting the detection.

Glove box supplier, glove box wholesaler, ) Methods:

① Close the cycle and analyzer



② Established the operating force to +five to + 8mbar

③Remove the h2o probe and speedily blind the air outlet with a blind plate

④Unscrew the metallic protective protect on the entrance of the probe, clean it with deionized h2o, and then dry it with dust-free of charge paper, then moisten the probe with dust-totally free paper dripped with cleansing fluid, and use evenly

⑤ Set it back in spot, open the cycle, turn on the analyzer following 1 hour, and return to regular stage right after fifty percent an hour.

notice:

Soon after the h2o probe is cleaned, the glove box stress is set to the unique value, and the cycle is turned on but the analyzer is turned on once more after 1 hour, due to the fact soon after the drinking water probe is cleaned, the probe floor is cleaned with deionized h2o or 65% concentrated phosphoric acid wetting contains h2o Sure, if the analyzer is turned on prematurely, the h2o worth will often be detected and displayed to exceed the regular, so it is needed to flow into in the tank for a while and wait around for the h2o vapor to dissipate before turning it on yet again.